TURBIDITY & SS

Principle: Near infrared LED (880nm) and 90° scattered light method in accordance with ISO 1027/EN 27027
Range: 0 to 500/4000NTU
0.00 to 1250mg/L, 50g/L
Resolution: 0.01 to 1NTU 0.01 to 1mg/L Unit NTU, FTU, ppm, mg/L, g/L
Accuracy: <±1%FS (Turbidity), <±5%FS (SS)
Repeatability: ±2%FS
Operate Temp: 32 to 122 °F (0 to 50°C)
Store Temp: 14 to 140 °F (-10 to 60°C)
Protection: Immersible, >IP68
Pressure: 5bar
Weight: 38.8Oz (1.1kg Sensor with 30’ cable)
Material: 316L, Sapphire Glass (Titanium Body is Optional)
Digital Output: Modbus RTU
Power: 24VDC from GDC
Consumption: At regular operation: 50mA(Max)
At cleaning operation: 240mA(Max)
Auto-Cleaning: Automatic Wiper cleaning system
CE Marking: EN 61326-1:2013 Electromagnetic Compatibility EMC